Fluke 714B Thermocouple Calibrator

$1,200.00

The Fluke 714B Thermocouple Calibrator is a dedicated temperature calibration instrument for measuring and simulating thermocouples and millivolt signals. It supports a broad range of thermocouple types for routine sensor and temperature transmitter calibration.

With 17 thermocouple types, 0–24 mA measurement, mV sourcing, linear ramp and automatic step functions, the 714B combines essential thermocouple calibration capabilities in a compact, rugged field instrument.

🕑 Lead time: In stock
Quantity:
Visa PayPal Mastercard American Express

Fluke 714B Thermocouple Calibrator

The Fluke 714B is a dedicated thermocouple calibrator designed for technicians who need to verify, simulate and calibrate thermocouple-based temperature instrumentation.

Unlike the broader Fluke 724 Temperature Calibrator, which also supports RTDs, resistance and voltage functions, the 714B focuses specifically on thermocouple and millivolt calibration. This focused design makes it a practical choice for applications where thermocouple calibration is the primary requirement.

Measure and Simulate 17 Thermocouple Types

The 714B can measure and source/simulate 17 thermocouple types, covering a wide range of industrial temperature measurement applications.

Supported thermocouple types include:


  • Type B

  • Type C

  • Type E

  • Type J

  • Type K

  • Type L

  • Type M

  • Type N

  • Type R

  • Type S

  • Type T

  • Type U

  • Additional thermocouple types supported by the instrument's calibration tables

It can also work directly with millivolt signals, making it useful for testing linear thermocouple transmitters.

Millivolt Source for Thermocouple Transmitters

Many temperature transmitters accept thermocouple signals and convert them into a standardized output such as 4–20 mA.

The 714B provides a -10 to 75 mV measurement and source range, allowing technicians to test the input circuit of thermocouple transmitters without physically heating or cooling a sensor.

This is particularly useful for:


  • Thermocouple transmitter calibration

  • Linear thermocouple transmitter testing

  • mV signal verification

  • Process temperature loop troubleshooting

The specified mV resolution is 0.01 mV.

Simultaneous Temperature and mA Measurement

The 714B can measure 4–20 mA while sourcing a temperature signal.

For example, during temperature transmitter calibration, the 714B can simulate a thermocouple input while measuring the transmitter's current output. This allows technicians to verify the relationship between the simulated temperature and the resulting 4–20 mA signal.

The current measurement specification is:


  • Range: 0–24 mA

  • Resolution: 0.001 mA

  • Accuracy: 0.010% + 2 μA

This makes the 714B useful not only as a thermocouple simulator but also for checking the electrical output of temperature transmitters.

Linearity Checks with Ramp and Step Functions

The Fluke 714B includes configurable source functions for checking transmitter linearity.

Available functions include:


  • Linear ramp

  • 25% step auto-ramp

  • Configurable source settings

  • Temperature source/simulation

  • mV source

These functions allow technicians to move through several temperature points automatically and observe the corresponding transmitter output.

For example:

0% → 25% → 50% → 75% → 100%

can be used to check whether the transmitter responds correctly across its calibrated range.

Compact Field Design

The 714B is designed as a portable, battery-powered field calibrator.

Important physical features include:


  • Backlit display

  • Magnetic hanging tool

  • Protective boot

  • Compact handheld housing

  • IP52 protection

  • 1 m drop specification

  • EMI-tolerant portable design

The included magnetic hanger allows technicians to attach the calibrator to suitable metal surfaces and keep both hands available during calibration work.

Calibration Accuracy and Traceability

The 714B provides both one-year and two-year accuracy specifications, depending on the thermocouple type and measurement range.

The instrument is supplied with a traceable calibration certificate with data, supporting applications where calibration traceability is required.

Key Specifications


  • Product type: Thermocouple calibrator

  • Thermocouple support: 17 types

  • Thermocouple function: Measure and source/simulate

  • mV range: -10 to 75 mV

  • mV resolution: 0.01 mV

  • mA measurement: 0–24 mA

  • mA resolution: 0.001 mA

  • mA accuracy: 0.010% + 2 μA

  • Linear ramp: Yes

  • 25% step auto-ramp: Yes

  • Display: Backlit

  • IP rating: IP52

  • Drop specification: 1 m

  • Operating temperature: -10 to +50 °C

  • Battery: 4 × AA

  • Dimensions: 52.5 × 84 × 188.5 mm

  • Weight: 515 g

  • Calibration certificate: Traceable certificate with data

  • Warranty: 3 years

Typical Applications

Thermocouple Calibration

Measure or simulate thermocouple signals to verify temperature sensors and associated instrumentation.

Temperature Transmitter Testing

Simulate a thermocouple input and measure the transmitter's 4–20 mA output.

Linear Thermocouple Transmitter Calibration

Use the mV source function to test transmitters that accept linearized thermocouple signals.

Process Temperature Loop Troubleshooting

Measure transmitter current while sourcing a controlled thermocouple signal.

Thermocouple Linearity Testing

Use ramp and step functions to verify transmitter response across multiple calibration points.

Field Instrument Maintenance

Use the compact battery-powered design for routine calibration and troubleshooting outside the laboratory.

FAQ

What is the Fluke 714B used for?

The Fluke 714B is used for measuring, sourcing and simulating thermocouple signals during temperature sensor and transmitter calibration.

How many thermocouple types does the Fluke 714B support?

The Fluke 714B supports 17 thermocouple types according to Fluke's current product specifications.

Can the Fluke 714B simulate a thermocouple?

Yes. It can both measure and source/simulate supported thermocouple signals.

Can the Fluke 714B test 4–20 mA temperature transmitters?

Yes. It can measure 0–24 mA while sourcing a temperature signal, allowing technicians to check the output of temperature transmitters.

Can the Fluke 714B source millivolts?

Yes. It provides a -10 to 75 mV measurement and source range for testing thermocouple transmitters and related instrumentation.

Does the Fluke 714B include a calibration certificate?

Yes. Fluke lists a traceable calibration certificate with data among the included items.

General Specifications

Maximum voltage applied between any terminal and earth ground or between any two terminals30 V
Operating temperature-10°C to 50°C
Storage temperature-30°C to 60°C
Operating altitude2,000 m
Storage altitude12,000 m
Relative humidity (% RH operating without condensation)Non condensing
90% (10°C to 30°C)
75% (30°C to 40°C)
45% (40°C to 50°C)
(Without condensation)
Vibration requirementsMIL-T-28800E, Class 2
Drop text requirements1 m
IP RatingIEC 60529: IP52
Electromagnetic environmentIEC 61326-1, Portable
SafetyIEC 61010-1, Max 30 V to earth, Pollution Degree 2
Power supply4 AA NEDA 1.5A IEC LR6 batteries
Size (H x W x L)52.5 x 84 x 188.5 mm
Weight515 g

DC mA Measurement

Range Resolution Accuracy (% of reading + counts)
0-24 mA0.001 mA0.010% + 2 μA
Temperature coefficient: ±(0.002% of reading + 0.002% of range) /°C (< 18°C or > 28°C)

Millivolt Measurement and Source

Range Resolution Accuracy (% of reading + counts)
-10 mV to 75 mV0.01 mV0.015% + 10 μA
Temperature coefficient: ±(0.002% of reading + 0.002% of range) /°C (< 18°C or > 28°C)

Thermocouple Input and Output

Accuracy (°C)

TC Type Range (°C) Measure (°C) Source (°C)
1 year 2 year 1 year 2 year
E-250 to 200°C1.32.00.60.9
-200 to -100°C0.50.80.30.4
-100 to 600°C0.30.40.30.4
600 to 1000°C0.40.60.20.3
N-200 to -100°C1.01.50.60.9
-100 to 900°C0.50.80.50.8
900 to 1300°C0.60.90.30.4
J-210 to -100°C0.60.90.30.4
-100 to 800°C0.30.40.20.3
800 to 1200°C0.50.80.30.3
K-200 to -100°C0.71.00.40.6
-100 to 400°C0.30.40.30.4
400 to 1200°C0.50.80.30.4
1200 to 1372°C0.71.00.30.4
T-250 to -200°C1.72.50.91.4
-200 to 0°C0.60.90.40.6
0 to 400°C0.30.40.30.4
B600 to 800°C1.32.01.01.5
800 to 1000°C1.01.50.81.2
1000 to 1820°C0.91.30.81.2
R-20 to 0°C2.32.81.21.8
0 to 100°C1.52.21.11.7
100 to 1767°C1.01.50.91.4
S-20 to 0°C2.32.81.21.8
0 to 200°C1.52.11.11.7
200 to 1400°C0.91.40.91.4
1400 to 1767°C1.11.71.01.5
C0 to 800°C0.60.90.60.9
800 to 1200°C0.81.20.71.0
1200 to 1800°C1.11.60.91.4
1800 to 2316°C2.03.01.32.0
L-200 to -100°C0.60.90.30.4
-100 to 800°C0.30.40.20.3
800 to 900°C0.50.80.20.3
U-200 to 0°C0.60.90.40.6
0 to 600°C0.30.40.30.4
BP0 to 1000°C1.01.50.40.6
1000 to 2000°C1.62.40.60.9
2000 to 2500°C2.03.00.81.2
XK-200 to 300°C0.20.30.20.5
300 to 800°C0.40.60.30.6
G100 to 300°C1.62.41.21.8
300 to 1500°C1.01.51.01.5
1500 to 2320°C2.03.01.62.4
D0 to 300°C1.62.41.21.8
300 to 1500°C1.01.51.01.5
1500 to 2315°C2.03.01.62.4
P0 to 1000°C1.62.40.60.9
1000 to 1395°C2.03.00.81.2
M-50 to 100°C1.01.50.40.6
100 to 1000°C1.62.40.60.9
1000 to 1410°C2.03.00.81.2