Temperature compensation: 0 °C to 50 °C (32 °F to 122 °F) to rated accuracy. For temperatures from -10 °C to 0 °C and 50 °C to 55 °C, add 0.005% FS/°C.
| Model | Range | Resolution | Accuracy | Burst Pressure |
|---|---|---|---|---|
| Fluke-700G01 | -10 to +10 inH2O (-20 to 20 mbar) |
0.001 inH2O (0.001 mbar) |
Positive ±0.1% range Vacuum ±0.1% range |
3 psi (210 mbar) |
| Fluke-700G02 | -1 to +1 psi (-70 to 70 mbar) |
0.0001 psi (0.001 mbar) |
Positive ±0.1% range Vacuum ±0.1% range |
5 psi (350 mbar) |
| Fluke-700G04 | -14 to 15 psi (-0.97 to 1 bar) |
0.001 psi (0.0001 bar) |
Positive ±0.05% FS Vacuum ±0.1% FS |
60 psi (4 bar) |
| Fluke-700G05 | -14 to 30 psi (-0.97 to 2 bar) |
0.001 psi (0.0001 bar) |
Positive ±0.05% FS Vacuum ±0.1% FS |
120 psi (8 bar) |
| Fluke-700G06 | -12 to 100 psi (-0.83 to 6.9 bar) |
0.01 psi (0.0001 bar) |
Positive ±0.05% FS Vacuum ±0.1% FS |
400 psi (26 bar) |
| Fluke-700G27 | -12 to 300 psi (-0.83 to 20 bar) |
0.01 psi (0.001 bar) |
Positive ±0.05% FS Vacuum ±0.1% FS |
1200 psi (80 bar) |
| Fluke-700G07 | -12 to 500 psi (-0.83 to 34 bar) |
0.01 psi (0.001 bar) |
Positive ±0.05% FS Vacuum ±0.1% FS |
2000 psi (138 bar) |
| Fluke-700G08 | -14 to 1000 psi (-0.97 to 69 bar) |
0.1 psi (0.001 bar) |
Positive ±0.05% FS Vacuum ±0.1% FS |
4000 psi (266 bar) |
| Fluke-700G10 | -14 to 2000 psi (-0.97 to 140 bar) |
0.1 psi (0.01 bar) |
Positive ±0.05% FS Vacuum ±0.1% FS |
8000 psi (550 bar) |
| Fluke-700G29 | -14 to 3000 psi (-0.97 to 200 bar) |
0.1 psi (0.01 bar) |
Positive ±0.05% FS Vacuum ±0.1% FS |
10000 psi (690 bar) |
| Fluke-700G30 | -14 to 5000 psi (-0.97 to 340 bar) |
0.1 psi (0.01 bar) |
Positive ±0.05% FS Vacuum ±0.1% FS |
10000 psi (690 bar) |
| Fluke-700G31 | -14 to 10000 psi (-0.97 to 690 bar) |
1 psi (0.01 bar) |
Positive ±0.05% FS Vacuum ±0.1% FS |
20000 psi (1035 bar) |
| Fluke-700GA4 | 0 to 15 psia (0 to 1 bar absolute) |
0.001 psi (0.0001 bar) |
±0.05% of range | 60 psi (4 bar) |
| Fluke-700GA5 | 0 to 30 psia (0 to 2 bar absolute) |
0.001 psi (0.0001 bar) |
±0.05% of range | 120 psi (8 bar) |
| Fluke-700GA6 | 0 to 100 psia (0 to 6.9 bar absolute) |
0.01 psi (0.0001 bar) |
±0.05% of range | 400 psi (27 bar) |
| Fluke-700GA27 | 0 to 300 psia (0 to 20 bar absolute) |
0.01 psi (0.001 bar) |
±0.05% of range | 1200 psi (80 bar) |
| Fluke-700RG05 | -14 to +30 psi (-0.97 to +2 bar) |
0.001 psi (0.0001 bar) |
Positive ±0.04% rdg + 0.01% FS Vacuum ±0.05% FS |
90 psi (6 bar) |
| Fluke-700RG06 | -12 to 100 psi (-0.83 to 6.9 bar) |
0.01 psi (0.0001 bar) |
Positive ±0.04% rdg + 0.01% FS Vacuum ±0.1% FS |
400 psi (27 bar) |
| Fluke-700RG07 | -12 to 500 psi (-0.83 to 34 bar) |
0.01 psi (0.001 bar) |
Positive ±0.04% rdg + 0.01% FS Vacuum ±0.1% FS |
2000 psi (138 bar) |
| Fluke-700RG08 | -14 to 1000 psi (-0.97 to 69 bar) |
0.1 psi (0.001 bar) |
Positive ±0.04% rdg + 0.01% FS Vacuum ±0.1% FS |
4000 psi (275 bar) |
| Fluke-700RG29 | -14 to 3000 psi (-0.97 to 200 bar) |
0.1 psi (0.01 bar) |
Positive ±0.04% rdg + 0.01% FS Vacuum ±0.1% FS |
10000 psi (690 bar) |
| Fluke-700RG30 | -14 to 5000 psi (-0.97 to 340 bar) |
0.1 psi (0.01 bar) |
Positive ±0.04% rdg + 0.01% FS Vacuum ±0.1% FS |
15000 psi (1035 bar) |
| Fluke-700RG31 | -14 to 10000 psi (-0.97 to 690 bar) |
1 psi (0.01 bar) |
Positive ±0.04% rdg + 0.01% FS Vacuum ±0.1% FS |
20000 psi (1380 bar) |
| 10 inH2O, 1, 15, 30 psi | Any clean dry non-corrosive gas |
| 100, 300, 500, 1000 psi | Any liquids or gases compatible with 316 stainless steel |
| Above 1000 psi | Any non-flammable, non-toxic, non-explosive, non-oxidizing liquid or gas compatible with 316 stainless steel |
| Dimensions (H x W x D) | 12.7 x 11.4 x 3.7 cm (5 x 4.5 x 1.5 in) |
| Pressure connection | 1/4 in NPT Male |
| Housing / Weight | Cast ZNAL; 0.56 kg (1.22 lb) with holster |
| Display | 5-1/2 digits, 16.53 mm (0.65 in) high; 20-segment bar graph, 0 to 100% |
| Power | Three (3) AA alkaline batteries |
| Battery life | 1,500 hours without backlight (continuous on); 2,000 hours at slow sample rate |