Endress+Hauser Memosens CCS51E Digital Free Chlorine Sensor

$560.00


The Endress+Hauser Memosens CCS51E is a digital free chlorine sensor for continuous HOCl measurement, featuring Memosens 2.0 technology, an amperometric measuring principle and measuring ranges up to 200 mg/l HOCl. It is designed for drinking water, process water, pool water and hygienic production applications where reliable disinfection control is required.

🕑 Lead time: 6-8 weeks
Quantity:
Visa PayPal Mastercard American Express

Reliable Free Chlorine Monitoring for Disinfection Control

The Memosens CCS51E is designed for applications where maintaining the correct free chlorine concentration is essential for effective disinfection.

Continuous monitoring helps operators maintain stable disinfectant levels, improve process control and avoid unnecessary chemical consumption.

Digital Memosens Technology for Easier Operation

The CCS51E uses Memosens 2.0 digital technology, storing sensor information and calibration data directly inside the sensor.

This allows easier sensor replacement, simplified commissioning and improved data integrity during operation.

Stable Amperometric Measurement

The sensor uses a closed, membrane-covered amperometric measuring cell to measure dissolved free chlorine.

The free chlorine is reduced at the cathode, generating a measurement signal proportional to the chlorine concentration. This proven principle provides reliable monitoring for continuous disinfection applications.

Designed for Water Treatment and Hygienic Applications

The CCS51E is suitable for various water qualities and disinfection processes, including:


  • Drinking water treatment

  • Process water monitoring

  • Food and beverage production water

  • Pool water disinfection

  • Utility water systems

It helps ensure effective disinfection while supporting safe and stable process operation.

Fast Response for Efficient Process Control

The sensor provides fast response times, including applications with low water volumes.

Combined with Liquiline transmitters, the CCS51E can be integrated into complete liquid analysis systems with additional parameters such as pH and ORP.

Key Features


  • Digital free chlorine sensor

  • Memosens 2.0 technology

  • Amperometric measurement principle

  • Measuring ranges:


    0–5 mg/l HOCl


    0–20 mg/l HOCl


    0–200 mg/l HOCl



  • Process temperature 0–55°C

  • Maximum process pressure 1 bar

  • Closed membrane-covered measuring cell

  • Two-electrode measuring system

  • Integrated temperature sensor

  • Fast response time

  • Compatible with Liquiline transmitters

  • Suitable for drinking water and process water applications

Typical Applications

Drinking Water Disinfection

Monitoring free chlorine levels to ensure reliable drinking water disinfection.

Food and Beverage Production

Checking disinfectant concentration in process water and supporting hygienic production conditions.

Pool Water Treatment

Helping optimize disinfectant dosing and maintain suitable chlorine levels.

Utility Water Systems

Monitoring free chlorine to protect processes and maintain water quality.

Frequently Asked Questions

What is the Endress+Hauser Memosens CCS51E used for?

The CCS51E is a digital sensor for measuring free chlorine (HOCl) concentration in water and process applications.

Is CCS51E a chlorine dioxide sensor?

No. The CCS51E measures free chlorine, while products such as CCS50E are designed for chlorine dioxide (ClO₂) measurement.

What measurement principle does CCS51E use?

The sensor uses an amperometric measurement principle with a closed membrane-covered measuring cell.

What is the measuring range of CCS51E?

Depending on the selected version, the measuring ranges are:


  • 0–5 mg/l HOCl

  • 0–20 mg/l HOCl

  • 0–200 mg/l HOCl

What industries use CCS51E?

Typical applications include drinking water treatment, food and beverage process water, pool water and industrial utility systems.

Which transmitter is used with CCS51E?

The CCS51E is designed for connection with compatible Endress+Hauser Liquiline transmitters using Memosens technology.

What are the advantages of Memosens technology?

Memosens stores calibration and sensor data digitally, helping simplify maintenance, improve data reliability and support predictive maintenance strategies.

Measuring Principle
Measuring principleFree chlorine
CharacteristicAmperometric measurement of dissolved free chlorine
Measuring methodClosed, membrane covered measuring cell
Reduction of free chlorine at the cathode
Application
ApplicationEnsuring reliable disinfection in drinking water
Process water
Dosing disinfectant efficiently in pool water
Detect the absence or presence of free chlorine in utilities
Guarantee food safety and provide hygienic packaging and bottling
Measuring Range
Trace0 to 5 mg/l HOCl
Standard0 to 20 mg/l HOCl
High0 to 200 mg/l HOCl
Design & Materials
DesignClosed amperometric 2-electrode measuring cell with PVDF membrane
Sensor shaftPOM
MembranePVDF
Membrane capPVDF
Sealing ringFKM
Diameter25 mm (0.98 inch)
Length161 mm (6.34 inch)
Process Conditions
Process temperature0 to 55 °C (32 to 130 °F), non-freezing
Process pressureMax. 1 bar (max. 14.5 psi)
Sensor & Connection
Temperature sensor10k NTC integrated (Memosens)
ConnectionInductive, digital connection head with Memosens